Clearing a Manual Tune Error Message in Analyst® Software


日期: 12/21/2017
类别: Analyst Software

0 投票数
   打印    评价文章:

For research use only. Not for use in diagnostic procedures.


Answer

If the syringe runs dry while doing direct infusion, a manual tune error may disrupt compound optimization in Analyst® software manual tune mode. If the error won't clear from the screen after refilling the syringe, try to deactivate the hardware profile and close the software. Restart Analyst services and re-launch Analyst software and the hardware profile. If the error message remains, the LC/MS system will need to be reset.