Resetting an Agilent Infinity II LC System after Receiving an ADD Error


日期: 10/31/2017
类别: Analyst Software

0 投票数
   打印    评价文章:

For research use only. Not for use in diagnostic procedures.


Answer

1. Deactivate the hardware profile in Analyst® software.

2. Close ADD and Analyst software.

3. Stop ADD and Analyst services (in this order)

4. Power off all modules.

5. Wait 30 seconds.

6. Power on all modules.

7. Open Analyst software (but do not activate the hardware profile).

8. In ADD, clear Configuration.

9. In ADD, use Auto-Configure to re-establish Configuration.

10. In ADD, check Status of modules, which should be green. It may take time (a few minutes) for the

modules to go to green.

11. Repeat above steps if the modules do not turn green.