Reporter 3.2: Including Only the Results for the Quantifier Ion in a Report


日期: 11/09/2019
类别: SCIEX OS , Analyst Software , MultiQuant Software

0 投票数
   打印    评价文章:

For research use only. Not for use in diagnostic procedures.


Answer

To include only the data associated with the quantifier ion in a report, rather than include information on both the qualifier and quantifier ions for a particular analyte, both of the transitions need to be grouped together in the quantitation method first. Then in a report template, use the For Each AnalyteGroup loop around the information to be reported (see example).
For the example below, only the calculated concentration for the quantifier ion will appear in the report output. User-added image