Analyst® Software Scheduled MRM Acquisition Failure Due To Unprogrammed Diverter Valve


日期: 03/14/2018
类别: Analyst Software

0 投票数
   打印    评价文章:

If a scheduled MRM (sMRM) acquisition methods created in Analyst® software fails to start, then it is very likely that the diverter valve doesn't have a time table programmed in the acquisition method. If the diverter valve is not required in your method, then right-click on the valve icon, unselect the Use option and save the method. Alternatively, the time table can be programmed for the divert valve to allow the sMRM method to acquire.