Analyst® TF Software Error: Instrument Optimization Parameters Are Not Available


日期: 12/13/2021
类别: Analyst Software

0 投票数
   打印    评价文章:

For research use only. Not for use in diagnostic procedures.


Issue Description

When running instrument optimization on a TripleTOF® system controlled through Analyst® TF software, the calibrant reference tables no longer appeared.


Resolution

This issue was attributed to the Windows operating system set to another language besides English (United States) in the settings. Once the language settings were updated to English (United States) under the regional time/date menu, normal operation of Analyst TF software resumed.