Error with MPX™ Software 1.2: Unable To Access Autosampler Database

When launching MPX™ software 1.2 on a Windows 7 64-bit operating system, an error appears, "Unable To Access Autosampler Database".

09/22/2023    MPX Software

Why Are My Run Times Longer after Upgrading to MPX Software 2.0?

Run times may be longer after upgrading to MPX software 2.0 because the acquisition methods may be different for the two streams. Make sure that the methods for both streams are the same.

11/02/2021    MPX Software

Is Cliquid® 3.4 Software Compatible with MPX™ 2.0 Software?

MPX™ Software is compatible with Cliquid® software and can be operated through the Cliquid software console; however, the Cliquid software Walk-up mode is not available.

04/16/2020    Cliquid Software | MPX Software

Location of the Event Logs for the MPX 2.0 Driver Software System

The event logs for the MPX 2.0 driver software system are saved under C:\Program Data\SCIEX\MPX Driver\Log\.

03/20/2020    MPX Software

Increasing the Maximum Injection Volume for an MPX™ Software System

To change the default injection volume in MPX™ 1.2.1 software, the MaxInjectVolume line in the MPX configuration file (Multiplex.INI) can be edited. Additionally, the value added to the injection volume field in the Analyst software batch editor (on the Sample tab) will override the configured MPX injection value during an acquisition run.

10/03/2019    MPX Software

Setting the Low Limit Value in the MPX 2.0 Software for Real-Time Data-Dependent Decision (RTD3) Flagging

When setting up MPX 2.0 software's carryover detection feature in the Define limit values screen (step 3), the low limit for each component must be defined using the lower limit of quantitation (LLOQ) or from the cut-off limit determined from the standard curve.

09/29/2019    MPX Software

How Very High Concentration Samples Are Handled in MPX 2.0 Software

MPX™ 2.0 software contains a feature for automated carryover monitoring, which can identify samples above a certain threshold.

09/26/2019    MPX Software

Setting the High Limit Value in the MPX 2.0 Software for Real Time Data-Dependent Decision (RTD3)

In the MPX software 2.0, the carryover detection feature requires that the values for the high or region height columns need to be defined for each component. These values are used to set the flag for the above upper concentration limit.

09/24/2019    MPX Software

MPX Softare Error: MPX Driver Is Not in the Active Analyst® Hardware Profile

To resolve the MPX software error "MPX driver is not in the active Analyst® hardware profile", remake the Analyst software hardware profile from scratch.

05/21/2018    MPX Software

MPX™ Software Batch Stoppage due to Daylight Savings Time Shift

If the computer clock updates for daylight savings time while MPX™ software is submitting a batch, data corruption and batch stoppage can occur. The clock must be set up so that automatic updates to the time do not occur.

03/12/2018    MPX Software

Where Are the MPX™ Driver Software Event Logs Located?

To locate the MPX™ Driver software event logs, navigate to the MPX Driver folder in Program Data and locate the Log folder.

10/31/2017    MPX Software

Re-installing a MPX License by Dropping the License Directly into the Program Files

If MPX driver software does not accept a license saved on the desktop during the license installation process, drop the license into the MPX bin folder in the program files.

03/31/2017    MPX Software