Issue Description

When submitting a batch in Analyst® software, an error message appears: Could not submit the batch data. DABServer Error=0x80040405 Unable to submit batch. You are exceeding the maximum number of waiting samples configured in the Queue Options.

image.png
 

Resolution

The solution is to increase the maximum number of waiting samples under the Queue Options menu in Analyst software. In Configure mode, go to Tools, Settings, Queue Options and increase the Max. Num. Waiting Samples to a number larger than number of samples in your batch. This should resolve the issue and the user will be able to submit samples.

An image of the queue options screen can be viewed in this article: https://sciex.com/support/knowledge-base-articles/analyst-how-to-increase-number-of-waiting-samples-in-the-queue_en_us.html