For research use only. Not for use in diagnostic procedures.
Answer
- In Analyst® Software go to the Quantitate Mode.
- Double click the Quantitation Wizard.
- Select the data file and the desired sample(s) to add to the new quantitation set.
- Select Create New Standard Query and type a Name.
- Specify the allowed accuracy variations (in percent). Max. Variation fields can be empty as desired.
- Specify which method will be used for this quantitation set or create a new method now.
- The Standard Query Status column shows if the allowed accuracy variation for QCs or STDs passed or failed.
- For a new results table, choose Select Existing and enable Execute Query as Standard Query to achieve the flagging for the Standard Query Status.
Posted: now